Load cell Data sheet

Loading...

Data sheet

Z6... Load cell

Special features - Welded‐on metal bellows - Nominal (rated) loads: 5 kg 0 1 t - Load cells and installation aids made of rust‐resistant materials - Verifiable up to 6000 divisions, test report as per OIML R60 - Six‐wire circuit - Optimized for parallel connection by off‐center load compensation - Meets EMC requirements as per DIN EN 45501 - Options: Ex‐protection designs as per ATEX 95

Dimensions (in mm; 1 mm = 0.03937 inches) Z6; Nominal (rated) loads 5 kg...500 kg

Z6; Nominal (rated) load 1 t

123

210 10

25 58

20

42 20

21

44

Metal bellows 13

15

23.4

34

9.9

27

31

Cable, 6‐wire, shielded, shield on enclosure

A

A 8.2 10.5

B1010-3.1 en

B 8.2 11.1

Metal bellows 11.1+0.1 38-0.1

Cable, 6‐wire, shielded, shield on enclosure

5...200 kg 500 kg

22"0.1 Load application

133"0.1

Load application

40

82 Mounting

70 40

Mounting

60

45 18

B

Cable 5.4; 3m long (standard version)

1

Specifications Type

Z6FD1

Accuracy class to OIML R 60 Number of scale intervals (nLC) Nominal (rated) load (Emax)

kg t % of Emax

Minimum scale division (vmin) Return of minimum dead load signal (DDR) Nominal (rated) sensitivity (Cn) Sensitivity tolerance with load appl. in spec. direction Temperature coefficient of sensitivity (TCS)2) Temperature coefficient of zero signal (TK0) Relative reversibility error (dhy)2) Non linearity (dlin)2) Load creep (dDR) in 30 min. Input resistance (RLC) Output resistance (R0) Reference voltage (Uref) Nominal supply voltage range (Bu) Insulation resistance (Ris) Nominal ambient temperature range (BT) Operating temperature range (Btu) Storage temperature range (Btl) Limit load (EL) Breaking load (Ed)

mV/V %

C3 3000 10; 20; 50; 100; 200; 500

1 0.0360 -

1 0.0090 -

Z6FC3MI C3/MI7.5 3000 50; 100; 200 0.5Emax/ 7500

Z6FC4

Z6FC6

C4 4000 20; 50; 100; 200; 500

C6 6000 50; 100; 200;

0.0066 -

-

2 "0.051)

+(1;-0.1) "0.0500 "0.0500 "0.0500 "0.0500 "0.0490

% of Cn/10 K % of Cn 

"0.0080 "0.0125 "0.0170 "0.0180 "0.0166

356 "0.2

V G oC

% of Emax

Nominal (rated) load Relative permissible oscillatory stress Nominal (rated) displacement (snom) approx. Weight, (G) approx. Degree of protection (IP) as per EN60529 (IEC529) Material:

Z6FC3

D1 1000 5; 10; 20; 50; 100; 200; 500

kg % of Emax mm kg

5 100 0.24 0.5

"0.0080 "0.0070 "0.0093 "0.0093 "0.0066 "0.0130 "0.0180 "0.0150 "0.0098 "0.0125 350...480 356 "0.12 5 0.5...12 u5 -10...+40 -30...+70 -50...+85 150 w300

"0.0040 "0.0093 "0.0080 "0.0110 "0.0083

10 20 50 100 200 500 100 100 100 100 100 70 0.3 0.29 0.27 0.31 0.39 0.6 0.5 0.5 0.5 0.5 0.5 0.5 IP 68 (tougher test conditions: 1 m water column;100 h)

1000 100 0.55 2.3

Stainless steel 3) Stainless steel 3) Stainless steel / Viton PVC

Measuring body Bellows Cable inlet gland Cable sheath

1)

For load cell Z6FC3/10kg: v $0.1 %. The values for linearity deviation, relative reversibility error and temperature effect on sensitivity are typical values. The sum of these values is within the cumulative error limits laid down by OIML R60. 3) As per EN 10088-1 2)

Options:

Ex‐protection designs as per ATEX 95:

II 2 G EEx ia IIC T4 or T6 (Zone 1) *) II 3 G EEx nA II T6 (Zone 2) II 2 D IP68 T80C (Zone 21) *) II 3 D IP68 T80C (Zone 22 for non‐conductive dust) *) with EC type examination certificate w w w w

Cable assignment (6‐wire configuration)

(gray) (black)

Sense lead (-) Excitation voltage (-)

With this cable assignment, the output voltage at the measuring amplifier is positive when the transducer is loaded.

(white)

Measurement signal (+)

(blue) (green)

Excitation voltage (+) Sense lead (+)

Measurement signal (-) (stranded Cable shield connected to enclosure ground wire) (red)

HBM

2

B1010-3.1 en

Installation aids, not included in scope of delivery (Dimensions (in mm; 1 mm = 0.03937 inches)) Note:

All installation aids are made of non‐rusting material. The rubber parts of the ZEL are made of chloroprene rubber.

Oscillating loading foot ZFP for nominal (rated) loads 5 kg...200 kg

69 ... 75

Height adjustable

Direction of load Z6/ZFP/200kg Steel Rubber  50  60

Knuckle eye ZGWR (maintenance‐free) for nominal (rated) loads 5 kg...1 t

Force feedback ZRR for nominal (rated) loads 5 kg ... 200 kg

Screw and washer are included in the scope of delivery K J

F1

H

SW

G1

M1 D

E1

L

M A

N

Z6/200kg/ZRR

a

F

G

Z6/.../ZGWR

E2

H W

D G2

B D

Nominal (rated) load

ZGWR

A

B

D

F

G

5...200 kg

Z6/200kg/ZGWR

16

8H7

24

36

500 kg

Z6/1t/ZGWR

20

10H7

28

43

28

43

M1

H F2

w

H

jJ

jK

L

M

SW

W

N

48

9

12.5

16

57

10.5

15

19

5

M8

14

12

46

6.5

M10

17

14

53

57

10.5

15

19

6.5

M10

17

14

55.5

1 t

Z6/1t/ZGWR

20

10H7

Nominal (rated) load

ZRR

D

E1

E2

F1

F2

G1

G2

H

M1

a

w

Depth

5...200 kg

Z6/200kg/ZRR

16

30

30

65

85

46

77

M8

M8x30

80 "1.1

123

15

Nominal (rated) load

Cone, conical pan ZK

jC

D

E

jU

X

5...200 kg

Z6/200kg/ZK

15

16

21

8.1-0.05

26

500 kg

Z6/1t/ZK

18

24

32

11-0.05

34

1 t

Z6/1t/ZK

18

24

32

11-0.05

36.5

B1010-3.1 en

3

HBM

Oscillating loading foot PCX for nominal (rated) loads 5 kg... 500 kg (Z6/PCX/500kg); 1 set comprising 4 pieces Z6/PCX/500kg Spacer

64 ... 70

Pos. 1, AF13 Pos. 2, AF17 Pos. 3, AF17

AF = Across Flats

10

69 ... 75

Angle: max. "5

Ø50 Ø60

Base plate / Mounting set for nominal (rated) loads 5 kg (Z6/ZPU/200kg) ... 500 kg (Z6/ZPU/500kg) Load application (Z6/...kg/ZPL; Z6/...kg/ZEL ; Z6/...kg/ZK)

View from below

82"0.1

15

14

66

31

Gap A

79

65

100

*

155

* Tightening torque MA: 23 Nm (200 kg); 45 Nm (500 kg) Gap A: In a load cell loaded with the nominal (rated) load, there should be a gap width 0.05 mm

22.5

200

Pendulum bearing ZPL for nominal (rated) loads 5 kg...1 t

E F

T

C M

U O

H

Z6/200kg/ZPL Z6/500kg/ZPL Z17/2t/ZPL (for 500 kg, large and 1 t nominal (rated) load

D

2 dust protection rings included in ZPL scope of delivery Fastening screw included in ZPL scope of delivery Nominal (rated) load

Pendulum bearing ZPL

jC

D

H

M

jO

T

E

F

jU

FR* (% of the load)

smax** (mm)

5...200 kg

Z6/200kg/ZPL

20-0.2

45

89+0.6 ‐0.8

M8

30

6.5

17

9

20D10

2.8

3.5

500 kg

Z6/500kg/ZPL

20-0.2

45

89+0.6 ‐0.8

M8

30

6.5

17

9

20D10

2.8

3.5

1 t

Z17/2t/ZPL

30-0.1

60

126.5

M10

46

8

22

14

30D10

2

7.5

* FR: Force feedback in N, with 1 mm lateral displacement ** smax: Maximum perm. lat. displacement with nominal (rated) loading

HBM

4

B1010-3.1 e n

Rubber‐metal bearing ZEL for nominal (rated) loads 5 kg...200 kg A

For nominal (rated) loads 1 t A

D

D

C

C

B

L

E F

H

F

E

H

Z6/1t/ZEL

Z6/200kg/ZEL

B

G

For nominal (rated) load 500 t

K

C

A

D

M N

F

H

P R

Z6/1t/ZEL

L

E

Screws offset by 35d

Correct installation position of the rubber‐metal bearing

B

Nominal (rated) load

ZEL

A

B

C

D

E

F

G

H

K

L

M

N

P

5...200 kg

Z6/200kg/ZEL

75

M12

12

40

79 "1.3

18.5

M8

SW17

19

-

500 kg

Z6/1t/ZEL

80

M10

10

39

105

+2.1 -2.2

26

-

SW27

-

20

-

-

120

100

1t

Z6/1t/ZEL

80

M10

10

39

117

+2.1 ‐2.2

26

-

SW27

-

20

120

100

smax**

R

FR*

-

-

163

3

9

60

400

4.5

9

60

400

4.5

* FR: Force feedback in N, with 1 mm lateral displacement ** smax: in mm, Maximum perm. lat. displacement with nominal (rated) loading

B1010-3.1 en

5

HBM

Hottinger Baldwin Messtechnik GmbH Im Tiefen See 45  64293 Darmstadt  Germany Tel. +49 6151 803‐0  Fax: +49 6151 803‐9100 Email: [email protected]  www.hbm.com

measure and predict with confidence

B1010-3.1 en

Subject to modifications. All product descriptions are for general information only. They are not to be understood as a guarantee of quality or durability.

Loading...

Load cell Data sheet

Data sheet Z6... Load cell Special features - Welded‐on metal bellows - Nominal (rated) loads: 5 kg 0 1 t - Load cells and installation aids made of...

411KB Sizes 48 Downloads 42 Views

Recommend Documents

Data sheet
start of lifetime: 66 W, end of lifetime: 66 W, reduction: 33 W, luminaire housing, of diecast aluminium, powder-coated,

Data sheet
Jan 24, 2013 - Acceptance standard. None; tolerances to ISO 9906. Class 2A / 3B; below 10 kW acc. to paragraph 4.4.2. Wa

DATA SHEET
April 2017. ITEM. PART NUMBERS. DESCRIPTION. QTY. 5CP6120. 5CP6120CSSG1. MATL. 283. 34334. —. Kit, Oil Drain (3/8” x

DATA SHEET
Application. The kWh meters types WQR96 MKII and WQ2R96 MKII are intended for energy measurement in single-phase and 3-

safety data sheet flow cell buffer - Nanopore Store - Oxford Nanopore
FLOW CELL BUFFER. According to Regulation (EU) No 453/ .... Protective equipment for fire-fighters. Self contained breat

Manual instruction - 4 load cell scales - Radwag
User Manual. 4 load cell platform scales. Manual number: ITKU-19-01-03-10-A. ▫ Low profiled scales. ▫ Mild steel sca

PIC16F84A Data Sheet - Microchip
High Performance RISC CPU Features: • Only 35 single word instructions to learn. • All instructions single-cycle exc

Major Works Data Sheet
Albert Camus was born in French colonial. Algiers. His father was killed in WWI, and he, his brother, mother, grandmothe

safety data sheet - Betco
Precautionary statements. Prevention. : Wear protective gloves: < 1 hour (breakthrough time): disposable vinyl. Wear eye

data sheet - Digital Fire
DESCRIPTION. HFC-227ea provides superior fire protection in a wide range of applications from sensitive electrical equip

Juicy Blunts Tube King Size for cones or tubes Rolling papers - 12 Pure FLAVORS | Primaire Antirouille Haute Resistance Rustol Cip 0.75L | Più cari